Physical
Weight: 10 Ounces (with batteries)
Size: 2.5 W x 4.5 H x 1.24 D inches (63.5 W 114.3 H x 31.5D mm)
Operating Temperature: -20° to 120° F (-30° to 50° C)
Case: Extruded aluminum body / nickel plated aluminum end caps (gasket sealed)
Keypad
sealed membrane that is resistant to both water and petroleum products.Nine tactilefeedback keys.
Transducer
Dual element (transmit and receive).
1 to 10 MHz frequency range.
Special high damped transducers available for use with through paint or coating measurements).
Locking quick disconnect LEMO connectors.
4 foot cable
Custom transducers available for special applications. |
Power Source
Two 1.5V alkaline or 1.2V NiCad AA cells
Typically operates for 200 hours on alkaline and 130 hours on NiCad.
Display flashes when battery is low. Unit turns off automatically when battery is too low to operate reliably.
Display
Multi-function 4.5 digit liquid crystal display with 0.500 inch numerals, backlit for use in poor light conditions.
Backlight is selectable:
on / off / auto ( illuminates only when taking a
measurement ).
Measurements displayed in inches, inches/microsecond. millimeters, and mters/second.
Bar graph indicates stability
of reading.
Data Logger (MMX-6DL Only)
Automatic numeric Data Logging system 1 to 1000 readings.
Certification
Factory calibration traceable to national standards.
Warranty>
5 year limited
|
Measuring
Range:
Pulse Echo Mode(Pit & Flaw Detection):
Measures from 0.025 to 19.999 inches (0.63 to
500 millimeters).
Echo Echo Mode (Thru Paint & Coatings):
Measures from 0.1 to 1 inch (2.54 to 25.4 millimeters). Extended range modification available on request: .2 to 4 inch (5 to 100 mm)
Switch to select English or metric units.
Resolution:
0.001 inch (0.01 millimeters)
Velocity Range:
.0492 to .3937 in/m s 1250 to 10,000 meters/sec
Built in:
Stainless steel reference disk for probe zeroing.
Four readings per second for single point measurements or 16 per second in Scan Mode – captures the minimum thickness.
Single and Two point calibration option included.
Features
Multi-Mode: Toggle between Pit/Flaw detection (Pulse Echo Mode), and through Paint/Coatings (Echo-Echo Mode) with the single press of a button.
Alarm Mode:
Enter a minimum acceptable thickness value. If measurement falls below minimum, red LED will illuminate and sound beeper. If measurement is above minimum value, green LED will illuminate.
Data Ouput: RS232 output sends thickness readings to a printer, computer or other external storage device. |