MP0
DUALSCOPE® MP0 Pocket-Size
Coating Thickness Gauge

by Fischer Technology


Easy, high-precision coating thickness measurements over ferrous & non-
ferrous metals.
Cutting-edge, pocket-size instruments with built-in probe.

The smallest coating thickness gauge can be the greatest for your measuring application.
 

The new DUALSCOPE® MP0 offers proven Fischer coating thickness probe technology. It measures coating thickness with high precision. It is user-friendly and sturdy. It is ideal for uncomplicated, reliable, on-site applications. With its integrated measurement probe, it measures the thicknesses of coatings quickly and non-destructively utilizing the magnetic induction as well as the eddy current method.

The DUALSCOPE® MP0 automatically recognizes the material type of the coated substrate material such as aluminum or steel, and selects the appropriate measurement method. Important statistical values are displayed via the statistics function.

With its superior repeatability, precision and standard-conforming calibration function, the DUALSCOPE® MP0 is unmatched in its class.

 
Features & Advantages

  • Measurements on all metallic materials (steel and iron or nonferrous metals).

  • Instrument automatically recognizes the substrate material under the coating and selects the appropriate measurement method.

  • Excellent repeatability and precision.

  • Low influence of permeability, electrical conductivity and geometry of the substrate material (curvature, thickness, etc.)

  • Patented conductivity compensation of the reading (eddy current method).

  • Small enough to make reliable measurements even in hard to reach areas.

Ideal for Daily Use

  • Immediate measurements without calibration.

  • One-point calibration for severe changes in shape or substrate material.

  • Easy-to-perform zero-point correction (normalization) for applications with little changes in shape and substrate material.

  • Statistics functions for display of mean value, standard deviation, MIN, MAX.

  • Quick, one-hand operation: Place instrument on specimen and see the measured value.

  • Two backlit LCD displays for readings in all instrument positions, even overhead.

  • Automatic On/Off function.

  • Durable, rugged housing and hard-material coated probe.

  • Low weight: only 132 g (4.7 oz) including batteries.

Additional Functions

  • Measurements with the "free-running display" mode for continuous scanning of surfaces, e.g., in tank construction.

  • Memory for 1000 readings.

  • Acoustic and optical measurement capture signal.

  • Units of measurement selectable in μm or mils.

  • Low Battery warning message

Applications & Specifications
 

Anodizing thickness on anodized aluminum.
 

Pain thickness measurement on iron.
 

Measurement of zinc on iron. A calibration is
mandatory for extremely curved specimens
to obtain correct measurements.
Integrated hard-material-coated probe with
V-notch in probe base for measurements
on cylindrical parts
 
Application Examples
  • Electroplated coatings made of zinc, chrome, copper on steel and iron.
  • Paint, lacquer, plastic coatings on steel and iron.
  • Paint, lacquer, plastic coatings on non-ferrous metals (aluminum, copper, brass) or on non-magnetic stainless steel.
  • Anodized coatings on aluminum

Specifications
 

Application NF, Iso/Fe Iso/NF

Thickness range

0-2000 µm

0-2000 µm

Accuracy based on Fischer standards

Up to 75 µm:

± 1.5 µm

75 to 1000 µm:

< 2%

1000 to 2000 µm:

< 3%

Up to 50 µm:

± 1 µm

50 to 1000 µm:

< 2%

1000 to 2000 µm:

< 3%

Repeatability & precision
based on Fischer standards

Up to 50 µm:

± 0.25 µm

50 to 2000 µm:

< 0.5%

Up to 100 µm:

± 0.5 µm

100 to 2000 µm:

< 0.5%

Measurement method

Magnetic induction
(DIN EN ISO 2178)

Eddy current
(DIN EN ISO 2360)

Operating temperature

5-60°C (41-140°F)

Data Memory

1000 measurements

Included with shipment

Instrument & manual

604-329

2 Batteries LR6., AA, 1.5 V

603-534

Calibration standard, Fe base

603-477

Calibration standard, Aluminum base

603-478

Calibration foil, 75 μm

603-479

 
* Abbreviations:
   NF = non-ferrous metals
   Fe = iron or steel
   Iso = isolating (i.e., paint)


 

Coating Inspection Instrumentation

 Cable & Metal Locators
 Dry Film Thickness
 Paint Inspection
 Wet Film Thickness
 Ultrasonic Thickness
 Holiday Detectors
 Surface Contamination
 Temperature & Humidity
 Surface Moisture
 Coating Adhesion
 Pit Depth Gauges
 Hardness & Inspection

 Coating Accessories

  

Corrosion Control Products Company

  

Corrosion Control Products Resources

 Search this Website
 Request Quote
 Order Products Online
 3-Ring Binder Catalog
 Industry Links
 CCP Suppliers
 Installation Guides
 MSDS
  

Farwest Cathodic
Protection Products
& Materials

  

Home

Products Search Request Quote Order Online

Corrosion Control Products Company
800-421-2623

Tel: 

 310-532-9314
Fax:  310-532-1472

1480 West Artesia Blvd., Gardena, CA 90248-3215, USA
 www.ccpco.com

© 1996-2009 Corrosion Control Products Company | Privacy Statement
 

Integrity – Service – Quality
                  
... Since 1956