Amprobe ACDC-3400 IND
CAT IV
Industrial True RMS Clamp Meter

True RMS Clamp-on multimeter for voltage to
750 VAC, 1000 VDC,
current to 1000 AAC,
plus ohms, frequency, capacitance & duty cycle.

Description
Amprobe’s ACDC-3400 IND Digital
Clamp-On Multimeter is CAT IV
600V / CAT III 1000V rated for industrial applications and
utilities. CAT IV rating is required for any application where the
electrical wiring is located outside of the building, connecting
electrical system to the power transformer. Includes True RMS (TRMS)
sensing for accuracy and dependability. Large jaw accommodates wires
up to 2" in diameter.
Features
- True RMS
- Measures AC Current up to 1000 ACA,
Voltage up to 750 VAC / 1000 VDC, Frequency, Resistance, Capacitance
and Duty Cycle.
- Peak Hold
- 1000A DC current
- Relative zero
- Audible continuity
- Auto and manual ranging
- Auto power off
- Data hold
- Backlight
- Diode Test
- Duty Cycle
- Bar-graph
- Accommodates conductors up to 2.00” (51
mm) in diameter
- Safety CAT IV 600 V, CAT III 1000 V
- Test leads, battery (installed),
Type-K thermocouple, users manual and carrying case included with
product.
- Wide choice of
Amprobe multimeter
accessories
|

|
DC Voltage Ranges |
660.0 mV, 6.600, 66.00, 660.0, 1000 V |
|
DC Voltage Accuracy |
± 0.5% rdg ± 2 digits |
|
AC TRMS Voltage Ranges |
660.0 mV, 6.600, 66.00, 660.0,
750 V (50-500 Hz) |
|
AC TRMS Voltage Accuracy |
± 1.5% rdg ± 8 digits |
|
AC TRMS
Input Impedance |
660 mV: >100 MΩ; 6.6 V:10 MΩ; 66 V to
1000 V: 9.1 MΩ |
|
Overload Protection |
1000 VDC or 750 VAC rms, 1000 AAC |
|
DC Current Ranges |
660.0A, 1000A |
|
DC Current Accuracy |
± 2.0% rdg ± 5 digits on
660A scale, ± 5.0% rdg ± 5 digits on 1000A scale |
|
AC TRMS Current Ranges |
660.0A, 1000A |
|
AC TRMS Current
Accuracy |
± 2.0% rdg ± 10 digits on
660A scale, ± 5.0% rdg ± 10 digits on 100A scale, 50/60
Hz. |
|
Overload Protection |
1000 A dc or 800A rms continuous |
|
True RMS Coupling |
5% to 100% of range |
|
True RMS Crest Factor |
< 3.0:1 |
|
Resistance Ranges |
660.0 Ω, 6.600, 66.00, 660.0 kΩ,
6.600 MΩ,
66.00 MΩ |
|
Temperature Ranges |
-35 to +1000°C (-30.0 to +1832°F) |
|
Temperature Sensor |
Type-K thermocouple |
|
Frequency (Auto-Ranging) |
66.00. 660.0 Hz, 6.600,
66.00, 660.0 kHz, 1.000 MHz |
|
Capacitance Ranges |
6.600, 66.00, 660.0 nF, 6.600 μF,
66.00 μF, 660.0 μF, 6.6 mF |
|
Diode Test |
1.0 mA, 3.2 Vdc |
|
Continuity Test |
Audible indication at < 30 Ω |
|
Duty Cycle Range |
5.0% to 95.0%, 40 Hz to 20
kHz |
|
Display |
3-3/4 digit (6600 count) LCD
display plus 66-segment analog bargraph |
|
Measurement Rate |
2.8 times per second |
|
Bargraph Rate Rate |
28 times per second |
|
Polarity |
Automatic |
|
Operating Temperature |
0°C to 50°C (32°F to 122°F
at < 70% RH |
|
Relative Humidity |
< 80% RH up to 31°C decreasing to 50% RH at
40°C |
|
Altitude |
Operation below 2000 m. Indoor use. |
|
Storage Temperature |
-20°C to 60°C, < 80% RH (with battery
removed) |
|
Power Supply |
Standard 9V battery, NEDA
1604, JIS 006P, IEC 6F22 |
|
Battery Life |
Typically 75 hrs with
carbon-zinc, 150 hrs with alkaline |
|
Max Conductor Diameter |
51 mm (2.00 in) |
|
Dimensions |
283 x 105 x 50 mm (11.1 x 4 x
2.1 in) |
|
Weight |
559 gm
(1.23 lb) |
|
Maximum Conductor |
50 mm (2.00 in) diameter |
|
Safety |
Meets EN61010-1:2001 and
EN61010-2-032:2002, CAT III 1000V, CAT IV 600V, class II
and pollution degree 2 |
|
Related Products |
Amprobe
ACDC-100 AC/DC Clamp-On Multimeter |

| Included Accessories

Carrying
case, instrument, test leads, Type-K thermocouple, user's manual.
Parts & Accessories
|
Optional Accessories |
TL1500 |
Cat IV 1000 V
(1500V Max) Test Leads with Alligator Clips |
|
Replacement Parts |
CC-ACDC |
Carrying Case |
|
|