Model B-3D
Dual-Channel Digital Multimeter
by MC Miller Co.


Advantages
- Increased versatility through solid state design
- Smaller and 30% lighter than Model B3A2
- Simplified operation
- RFI shielded throughout
- Excellent visibility in bright sunlight
- Selectable input resistance
- Large liquid crystal digital display
General Information
- Right Meter: Selectable Input Resistances
of 10, 25, 50,100 & 200 megohms on all ranges from 200 mV to 200 V; 20 mV range fixed
at 1000 ohms.
- Left Meter: 10 Megohms Input Resistance on
all ranges from 200 mV to 20 V; 20 mV range fixed at 1000 ohms.
- RFI Rejection: A major design achievement
was to eliminate the effects of RFI, and offer accurate, repeatable readings.
- Logic: C-MOS LSI; crystal controlled
timing.
- AC Rejection: Normal mode 50/60/400 Hz
greater than 60 dB.
- Zero: Automatic.
- Polarity: Automatic (negative symbol
displayed, positive assumed).
- Decimal Point: Automatic.
- Battery Life: Amplifiers approximately 150 hours
continuous. Low battery voltage indication incorporated into display. Other batteries have
life dependent upon use.
- Battery Types: Only standard batteries
used: D-size, penlight & 9 V transistor types.
- Accuracy: DC 1% of reading
+ 1
digit; AC 2% reading + 1 digit.
|

| Specifications
Right Meter
- Liquid Crystal Display with five DC ranges
from 20 mV to
200 V.
- High input resistances of 10, 25, 50, 100 & 200 megohms
selectable by rotary switch, applicable to 200 mV, 2 V and 200 V ranges. High input
resistance permits accurate voltage measurement if there is high resistance in the
external circuit. This is an important factor in structure to earth potentials, where
reading errors can be costly. The 20 mV range has a fixed input resistance of 1000 ohms.
- One AC range 200 V.
Left Meter
- Liquid Crystal Display with 4 ranges from 20 mV to 20 V.
- Milliammeter/ammeter with 4 ranges from 20 mA to 20 A (20 mV
drop shunts).
- Three direct reading ohmmeter ranges: 20 ohms, 2 kOhms, 2
MOhms.
DC Bias
- A DC Bias circuit is connected in series with the right
meter and can be used to balance out galvanic and earth potentials.
- A biasing potential of full scale (+ or -) on any range
permits direct measurements of either positive or negative small changes in potential.
- A panel-mounted light is activated as a reminder if the DC
Bias circuit is turned on.
Ammeter
- Left meter can serve as a milliammeter / ammeter with
built-in shunts to provide a wide selection of ranges.
- Toggle switch permits the left meter to be used as a
conventional ammeter, a zero resistance ammeter, or to connect the rheostat and controls
in to the circuit to adjust current to a desired value for various tests. Current from a
few milliamperes to about 5 amperes, supplied from either internal batteries or an
external source, can be adjusted with the controls. Used for current requirement tests,
soil resistivity tests, etc.
- A second toggle switch connects the right meter to the left
hand terminals for resistance testing. The left meter can be used with batteries and
controls to measure current while the right meter is being used to measure IR drop. This
facilitates making ''Null-Amp" type current measurements.
Ohmmeter
- Three built-in ohms ranges.
- Used to determine when good low-resistance contacts are
obtained.
- Circuit can also be used for checking continuity of test
leads, resistance of bond wires, etc.
Overload Protection
- Advanced design features virtually eliminate the chance of
damage to either meter from electrical overload on all volt and millivolt ranges, even the
20 mV range. Tests indicate that the B-3D will withstand AC or DC voltages far in excess
of any encountered in normal field testing.
- Ammeter shunts are not protected by the amplifier and are
subject to damage by current overloads.
Displays
- Liquid crystal displays are special high temperature type
(18oF to 176oF or 14oC to 80oC) built to rigid
M.C. Miller Co. specifications.
- Large display characters (3/4") facilitate rapid,
accurate readings.
- Built-in annunciators to reduce errors in recording
readings: V, MV, A, MA, Ohms, LoBat, AC.
- Excellent visibility in bright sunlight.
- Damaged displays can be economically replaced in the field
without special tools.
Shunts
- Built-in multi-range ammeter shunt provide high overload
capacity and a high degree of accuracy.
- Optional accessory external shunt can extend range of
current measurements to 100 amperes.

|
|